mr2.algorithms.reconstruction.GrappaReconstruction

class mr2.algorithms.reconstruction.GrappaReconstruction[source]

Bases: Reconstruction

Calibrate GRAPPA, interpolate Cartesian k-space and reconstruct coil images.

__init__(kdata: KData, acs: Tensor, kernel_size: tuple[int, int, int], *, regularization: float = 0.01, csm: Callable[[IData], CsmData] | CsmData | SensitivityOp | None = CsmData.from_idata_walsh) → None[source]

Initialize the reconstruction and calibrate reusable kernels.

Parameters:
  • kdata (KData) – Cartesian acquisition defining the encoding grid and sampling pattern. All batches must share the same pattern. Samples must not repeat.

  • acs (Tensor) – Fully sampled Cartesian calibration tensor (..., coils, kz, ky, kx). Must use the same coil basis and frequency spacing as kdata; its batch dimensions broadcast to the acquired data.

  • kernel_size (tuple[int, int, int]) – Positive odd neighborhood widths in z, y, x. Use a z width of one for 2D acquisitions. At boundaries, only in-grid acquired sources enter the fit. Every missing point needs an acquired neighbor.

  • regularization (float, default: 0.01) – Relative ridge strength passed to calibrate.

  • csm (Callable[[IData], CsmData] | CsmData | SensitivityOp | None, default: CsmData.from_idata_walsh) – Coil combination maps, operator or calculation from completed coil images, following DirectReconstruction. Set to None to return separate coil images. For differentiable reconstruction, construct on the target device.

acs: Tensor

Calibration data in the acquisition coil basis.

kernels: Tensor

Fitted kernels indexed by source pattern.

dcf_op: DensityCompensationOp | None

Density Compensation Operator.

csm_op: SensitivityOp | None

Coil Sensitivity Operator.

noise: KNoise | None

Noise Data used for prewhitening.

fourier_op: LinearOperator

Fourier Operator.

__call__(kdata: KData) → IData[source]

Apply the reconstruction.

forward(kdata: KData) → IData[source]

Interpolate and reconstruct data with the calibrated sampling pattern.

Parameters:

kdata (KData) – Data with the trajectory and coil basis used during initialization.

Returns:

Reconstructed image data, optionally combined using coil sensitivities.

static calibrate(acs: Tensor, source_pattern: Tensor, *, regularization: float = 0.01) → Tensor[source]

Fit a GRAPPA kernel from fully sampled Cartesian calibration data.

Parameters:
  • acs (Tensor) – Fully sampled Cartesian ACS, shape (..., coils, kz, ky, kx). Leading batch dimensions are calibrated independently.

  • source_pattern (Tensor) – Boolean 3D neighborhood with positive odd widths. True selects an acquired source offset; the center is the target and must be False. Only complete ACS windows are used.

  • regularization (float, default: 0.01) – Nonnegative ridge strength, relative to the mean diagonal of S.H @ S for source matrix S. Zero uses a minimum-norm pseudoinverse. Batches with zero source energy return zero kernels.

Returns:

Complex kernel with shape (..., coils_out, coils_in, *kernel_size). Unused offsets are zero.

interpolate(kdata: KData) → Tensor[source]

Fill the encoded Cartesian grid, preserving acquired samples exactly.

Parameters:

kdata (KData) – Data with the trajectory and coil basis used during initialization.

Returns:

Completed k-space tensor, shape (..., coils, kz, ky, kx).

recalculate_csm(kdata: KData, csm_calculation: Callable[[IData], CsmData] = CsmData.from_idata_walsh, noise: KNoise | Literal[False] | None = None) → Self[source]

Update coil sensitivities from GRAPPA-completed images.

Parameters:
  • kdata (KData) – Acquisition with the trajectory and coil basis used for calibration.

  • csm_calculation (Callable[[IData], CsmData], default: CsmData.from_idata_walsh) – Sensitivity calculation receiving the completed coil images.

  • noise (Union[KNoise, Literal[False], None], default: None) – Must be None or False. For prewhitening, transform acquisition and ACS together before calibrating GRAPPA kernels.

recalculate_fourierop(kdata: KData) → Self[source]

Update the Cartesian sampling operators and recalibrate from stored ACS.

Parameters:

kdata (KData) – Acquisition defining the new encoding grid and sampling pattern.

direct_reconstruction(kdata: KData, *, csm: CsmData | Literal[False] | None = None, noise: KNoise | Literal[False] | None = None) → IData[source]

Direct reconstruction of the MR acquisition.

Here we use \(S^H F^H W\) to calculate the image data using the coil sensitivity operator \(S\), the Fourier operator \(F\), and the density compensation operator \(W\).

Parameters:
  • kdata (KData) – k-space data

  • csm (Union[CsmData, Literal[False], None], default: None) – Coil sensitivity maps used for coil combination. If None, self.csm_op is used. If False, no coil combination is performed.

  • noise (Union[KNoise, Literal[False], None], default: None) – Noise measurement for prewhitening. If None, self.noise is used. If False, no prewhitening is performed.

Returns:

image data