mr2.phantoms.a4im_lowfield.FreeMaxDataset

class mr2.phantoms.a4im_lowfield.FreeMaxDataset[source]

Bases: IsmrmrdDataset

Raw knee and phantom data from a Siemens MAGNETOM Free.Max at 0.55 T.

The 80 cm-bore DryCool superconducting system provides 22–26 mT/m gradients and 45–55 T/m/s slew rate. Twenty volunteers were scanned bilaterally with a flexible six-element Contour S coil. The raw subject files nevertheless report 15 active receive channels (some phantom files report 6). The sagittal protocol contains 2D Cartesian TSE PDw/T1w/T2w scans (TE/TR 28/3070, 12/490, and 78/3040 ms; ETL 9/4/12; 112 x 112 matrix; 10–14 slices) and 2D MOLLI (3-3-5) T1 and T2-prepared bSSFP mapping (11 and 6 sets, 128 matrix, 1.6 x 1.6 x 8 mm, six slices). The descriptor reports 0.8 x 0.8 x 5 mm for the TSE data, whereas the raw reconstruction metadata imply about 1.52 x 1.52 x 5 mm. All acquisitions have one average.

Each of the 20 subjects and 3 phantoms contains 16 files. The 6 qualitative scans intended for vendor Deep Resolve processing are excluded, leaving 10 raw files per subject or phantom. The published Subject9/raw/pd_tse_sag_Right.mrd file is truncated and is excluded with a warning. See Zenodo record 20516472.

__init__(path: str | PathLike | Sequence[str | PathLike] = CACHE_DIR_A4IM_LOWFIELD / 'free_max', acquisition: Literal['pdw', 't1w', 't2w', 't1map', 't2map'] | None = None, *, trajectory: KTrajectoryCalculator | KTrajectory | KTrajectoryIsmrmrd | None = None) → None[source]

Select Free.Max files.

Parameters:
load(filename: str | PathLike) → KData[source]

Load an ISMRMRD file using this dataset’s loading options.

__eq__(value, /)

Return self==value.

__new__(**kwargs)